01
Front

A calibration line is A=0.025c+0.020A=0.025c+0.020. What is cc when A=0.520A=0.520?

Back

Solving the fitted line gives c=0.520−0.0200.025=20.0c=\frac{0.520-0.020}{0.025}=20.0 mg/L.

02
Front

What belongs in a spectrophotometric blank?

Back

A blank contains the solvent and relevant reagents or matrix components, but no analyte. It establishes the baseline and compensates for background absorption, including absorption by the cuvette.

03
Front

What absorbance corresponds to transmittance T=0.10T=0.10?

Back

Since A=−log⁡10(T)A=-\log_{10}(T), A=−log⁡10(0.10)=1.00A=-\log_{10}(0.10)=1.00. Absorbance is dimensionless.

04
Front

What guides the choice of an analytical wavelength?

Back

Choose a wavelength that gives the analyte a useful signal while minimizing contributions from the blank and likely interferents; also consider robustness to wavelength-setting error.

05
Front

State the Beer–Lambert relationship and identify its variables.

Back

For one absorbing species under suitable conditions, A=εbcA=\varepsilon bc, where ε\varepsilon is molar absorptivity, bb is path length, and cc is analyte concentration.

06
Front

What does an absorption spectrum show, and what do its peaks indicate?

Back

An absorption spectrum plots absorbance against wavelength. Its peaks mark wavelengths of strong absorption; peak position and shape can help characterize an absorbing species.

07
Front

Why is λmax⁡\lambda_{\max} often selected for measurement?

Back

The wavelength of maximum absorbance, λmax⁡\lambda_{\max}, often provides a strong signal and is relatively insensitive to small wavelength-setting errors near the peak.

08
Front

Under what fixed conditions is absorbance proportional to concentration?

Back

At fixed wavelength and path length, absorbance is proportional to concentration over a suitable verified linear range.

09
Front

How are standards arranged on a linear calibration plot?

Back

Measure standards spanning the expected concentration range, plot absorbance AA on the vertical axis against concentration cc on the horizontal axis, and fit an appropriate line.

10
Front

Why might a solvent-only blank be inadequate?

Back

A solvent-only blank may not correct for absorption by reagents or sample-matrix components. Use a blank that matches the standards and unknown as closely as possible.

11
Front

What should you do if an unknown falls outside the validated calibration range?

Back

Do not extrapolate beyond the validated range. Dilute or otherwise prepare the unknown so its measurement falls within the range supported by the standards and observed linearity.

12
Front

What does higher absorbance tell you about transmitted light?

Back

Higher absorbance means less light is transmitted through the sample.