Spectrophotometry: Measurement and Calibration Online Quiz Questions

Use this free practice quiz with 10 questions to review Spectrophotometry: Measurement and Calibration, test your knowledge, and prepare for your next test or exam.

10 questions
01
Choose one
1 point

When selecting a measurement wavelength for an analyte in a sample that may contain interferents, which approach is most appropriate?

  1. A

    Always use the wavelength with the highest peak, regardless of other absorbing substances.

  2. B

    Choose a wavelength that provides a useful analyte signal while limiting interference from the sample.

  3. C

    Choose the wavelength where the blank absorbs most strongly.

  4. D

    Avoid the absorption maximum because wavelength-setting errors are always greatest there.

02
Choose one
1 point

Which procedure correctly describes using a linear calibration curve to determine an unknown concentration?

  1. A

    Plot concentration on the vertical axis and absorbance on the horizontal axis, then use the curve only outside the standards’ range.

  2. B

    Measure only the unknown and infer its concentration from the highest standard.

  3. C

    Plot absorbance against concentration for standards, fit a line, and use it to find an unknown within the validated range.

  4. D

    Use the fitted line to determine concentration without measuring standards.

03
Choose one
1 point

A sample matrix contains a reagent that absorbs light. Which blank is most appropriate for correcting the background when measuring an analyte in that matrix?

  1. A

    Use a blank with the same solvent and relevant reagents or matrix components as the samples, but without the analyte.

  2. B

    Use a blank containing the analyte at the same concentration as the unknown.

  3. C

    Use only pure solvent even when the samples contain absorbing reagents.

  4. D

    Use a blank containing the analyte but no solvent.

04
Choose one
1 point

For a single absorbing species under suitable linear-range conditions, what happens to absorbance if concentration doubles while wavelength and path length remain fixed?

  1. A

    The absorbance is halved.

  2. B

    The absorbance stays the same.

  3. C

    The absorbance becomes four times as large.

  4. D

    The absorbance doubles.

05
Choose one
1 point

Which statement best describes an absorption spectrum and how to interpret its peaks?

  1. A

    It plots transmittance against concentration, and each peak uniquely identifies the analyte.

  2. B

    It plots absorbance against wavelength; peaks indicate strong absorption, though other absorbing substances may contribute.

  3. C

    It plots wavelength against path length, and peaks indicate the sample’s concentration.

  4. D

    It plots concentration against absorbance, and peaks show wavelengths where no light is absorbed.

06
Choose one
1 point

Two candidate wavelengths provide useful analyte signals. One is on a broad, relatively flat peak and the other is on a very sharp peak. Which statement about robustness to small wavelength-setting errors is supported?

  1. A

    A very sharp peak is always more robust because its absorbance changes rapidly with wavelength.

  2. B

    A broad peak is unsuitable because it cannot provide a useful signal.

  3. C

    A broad, relatively flat peak may be more robust to small wavelength-setting errors than a sharp peak.

  4. D

    Peak shape has no bearing on sensitivity to wavelength-setting errors.

07
Choose one
1 point

The analyte has its strongest peak at one wavelength, but a likely interferent also absorbs strongly there. What is a reasonable wavelength-selection decision?

  1. A

    A wavelength away from the maximum may be better if it reduces interference while retaining a useful analyte signal.

  2. B

    The maximum must always be chosen, even if an interferent absorbs strongly there.

  3. C

    The wavelength where the blank absorbs most is best because it maximizes total signal.

  4. D

    Choose a wavelength where the analyte has no absorbance so interferents cannot affect the result.

08
Choose one
1 point

A validated calibration line is A=0.025c+0.020A = 0.025c + 0.020, where cc is in mg/L. A diluted unknown has absorbance 0.5200.520, and the dilution factor is 4. What was the unknown’s original concentration?

  1. A

    5.0 mg/L

  2. B

    20.0 mg/L

  3. C

    24.0 mg/L

  4. D

    80.0 mg/L

09
Choose one
1 point

Why might a solvent-only blank be inadequate for standards and unknowns that contain absorbing reagents or other matrix components?

  1. A

    It removes the analyte from the standards and unknown.

  2. B

    It may fail to correct for absorbing reagents or sample-matrix effects.

  3. C

    It automatically corrects for every possible instrument effect.

  4. D

    It makes the calibration curve independent of the blank.

10
Choose one
1 point

Which equation expresses the Beer–Lambert relationship among absorbance AA, molar absorptivity ε\varepsilon, path length bb, and concentration cc?

  1. A

    A=ε+b+cA = \varepsilon + b + c

  2. B

    A=εbcA = \frac{\varepsilon}{bc}

  3. C

    A=εbcA = \varepsilon b c

  4. D

    A=bcεA = \frac{b c}{\varepsilon}