Interpreting Spectrophotometry Data

A practical guide to interpreting spectra and obtaining reliable concentrations from spectrophotometric measurements.

and the

and the

, TT, is the ratio of transmitted light intensity, II, to incident intensity, I0I_0:

T=II0,A=−log⁡10(T)=log⁡10(I0I).T=\frac{I}{I_0}, \qquad A=-\log_{10}(T)=\log_{10}\left(\frac{I_0}{I}\right).

Thus, lower means higher . For example, if T=0.10T=0.10, then A=−log⁡10(0.10)=1.00A=-\log_{10}(0.10)=1.00: one tenth of the incident light is transmitted. has no units.

An plots against wavelength. Its peaks show where the sample absorbs strongly, and a peak’s position and shape can help characterize an absorbing species. They do not necessarily identify a substance on their own: other absorbing components in the sample may also contribute.

Choosing a useful wavelength

Choosing a useful wavelength

First record or consult the analyte’s under the conditions you will use. The wavelength of maximum , λmax⁡\lambda_{\max}, is often a good starting point: it provides a strong analyte signal and, near the peak, small wavelength-setting errors may have relatively little effect on .

But the tallest peak is not always the best choice. Other sample components may absorb at that wavelength, so another wavelength could reduce interference. Peak shape matters too: a broad, relatively flat peak may be less sensitive to wavelength-setting errors than a very sharp one. Choose a wavelength that gives the analyte a useful signal while keeping contributions from the and likely interferents as small as possible.

Correcting background and relating to concentration

Correcting background and relating to concentration

A contains the solvent and, when relevant, the reagents and other matrix components—but not the analyte. Measuring it establishes a baseline and compensates for light absorbed by those components and by the cuvette. Use a that matches the standards and unknown as closely as possible. A solvent-only may leave background from absorbing reagents or the sample matrix uncorrected. An unexpectedly high or a substantial calibration intercept can signal a mismatched , a contaminated cuvette, or another background contribution.

For a single absorbing species under suitable conditions, the is

A=εbc,A=\varepsilon bc,

where ε\varepsilon is the molar absorptivity at the selected wavelength, bb is the light path length, and cc is the analyte concentration. At fixed wavelength and path length, is proportional to concentration: increasing concentration increases within the applicable linear range. This relationship is not guaranteed to remain linear in every sample. High concentrations, light scattering, chemical changes, or instrumental effects can cause deviations.

Calibrating and measuring an unknown

Calibrating and measuring an unknown

A connects measured to known concentrations. First the instrument with an appropriate , then measure standards spanning the expected concentration range at the selected wavelength. Plot , AA, on the vertical axis and concentration, cc, on the horizontal axis, and fit an appropriate line. For a linear calibration,

A=mc+b0,A=mc+b_0,

where mm is the slope and b0b_0 is the intercept. Measure the unknown’s and rearrange the fitted equation to solve for concentration:

c=A−b0m.c=\frac{A-b_0}{m}.

For the source example, the fitted line is A=0.025c+0.020A=0.025c+0.020, with cc in mg/L. If the unknown has A=0.520A=0.520, subtract the intercept and divide by the slope:

c=0.520−0.0200.025=20.0 mg/L.c=\frac{0.520-0.020}{0.025}=20.0\text{ mg/L}.

This is the concentration of the measured solution. If the sample was diluted before measurement, multiply that result by the to obtain the original concentration.

Use the calibration only within the range supported by the standards and observed linearity; do not extrapolate beyond it. If an unknown falls outside that range, dilute or otherwise prepare it so its measurement falls within the validated range. A good-looking fit statistic alone does not show that standards, , or sample were prepared correctly.

  • : The wavelength at which reaches a maximum.

  • : A sample component other than the analyte that can contribute to the measured signal.