2 Blood Pressure Assessment

Learn what blood pressure measures, how it is determined and measured, how to interpret adult readings, and how to recognize signs of worsening perfusion.

What represents

(BP) is the force blood exerts against arterial walls. It is recorded in millimeters of mercury as systolic pressure over diastolic pressure.

  • (SBP) is the peak arterial pressure during ventricular contraction.

  • (DBP) is the lowest arterial pressure between contractions.

  • (PP) is the difference between systolic and diastolic pressure: PP=SBP−DBPPP = SBP - DBP. It is influenced by and arterial stiffness.

  • (MAP) is the approximate average arterial pressure over one cardiac cycle. At ordinary heart rates, a useful estimate is:

MAP≈SBP+2×DBP3MAP \approx \frac{SBP + 2 \times DBP}{3}

For a pressure of 120/80 mm Hg120/80\ \text{mm Hg}, estimated MAP is about 93 mm Hg93\ \text{mm Hg}. This estimate is less accurate at unusually high heart rates or when the arterial waveform is abnormal. MAP is a useful indicator of perfusion pressure, but it does not by itself prove that organs are adequately perfused.

How is determined

A simplified relationship connects with and :

MAP≈CO×SVRMAP \approx CO \times SVR

is the volume of blood pumped by a ventricle per minute. It depends on heart rate and :

CO=HR×SVCO = HR \times SV

More precisely, the pressure gradient driving systemic flow also reflects venous pressure. The simplified relationship is useful for understanding how changes in flow and resistance affect arterial pressure.

is affected by , contractility, and . describes ventricular filling and myocardial stretch at the end of diastole; venous return and circulating blood volume influence it. Low volume can reduce and , while excessive volume may cause congestion.

is the resistance or pressure a ventricle must overcome to eject blood. Increased arterial resistance or pressure can raise and, in some circumstances, reduce . is strongly influenced by small-artery and arteriole diameter: vasoconstriction generally increases resistance, while vasodilation decreases it.

These factors interact. Vasodilation can result in low despite normal or increased . With blood loss, a person may initially maintain through compensatory vasoconstriction and a faster heart rate.

Cuff measurement and technique

A manual measurement uses an inflatable cuff and stethoscope. As cuff pressure is released, the first Korotkoff sound indicates systolic pressure; in adults, disappearance of the sounds indicates diastolic pressure. Validated automated upper-arm devices use an oscillometric method to estimate from cuff-pressure oscillations.

Use a properly maintained, validated device and a cuff that fits the measured arm. A cuff that is too small can produce a falsely high reading; one that is too large can produce a falsely low reading.

Reliable seated measurements

  1. Have the person rest quietly for at least 55 minutes. They should avoid talking or moving during the reading.

  2. Seat the person with their back supported, feet flat on the floor, and legs uncrossed.

  3. Place the correctly sized cuff on a bare upper arm, and support the arm at heart level.

  4. Take two readings at least 11 minute apart. Record the readings and which arm was used.

  5. For ongoing home monitoring, use a validated upper-arm monitor and follow a consistent schedule. Repeated readings are more informative than one isolated value.

Recent activity, caffeine, nicotine, a full bladder, unsupported posture, talking, and an incorrectly sized cuff can affect results. If a reading is unexpected, check the technique and repeat it when clinically appropriate rather than relying on a single number.

Other monitoring methods

involve measuring and pulse after the person rests supine and again after standing, according to the applicable clinical protocol. Record the person's position, timing, symptoms, and readings. Dizziness or faintness on standing may indicate impaired compensation or reduced circulating volume.

uses an automated cuff to take repeated measurements during usual daily activities and sleep. It can help characterize outside the clinic.

An provides continuous, beat-to-beat pressure measurement and may permit arterial blood sampling. Its fluid-filled system's transducer must be correctly leveled and zeroed, and the waveform checked. If the transducer is too high, the displayed pressure may be falsely low; if it is too low, the displayed pressure may be falsely high.

Other hemodynamic measurements may help assess circulation in critically ill patients. Central venous pressure and pulmonary artery occlusion pressure are pressure measurements used in context to estimate aspects of right- and left-sided filling. They are not direct measurements of blood volume or a stand-alone measure of fluid responsiveness. Interpret them alongside the clinical picture and other findings.

Interpreting adult readings

For adults, the 2025 AHA/ACC guideline categorizes office using the average of at least two careful readings obtained on at least two occasions. When systolic and diastolic readings fall into different categories, use the higher category.

  • Normal: systolic pressure is below 120 mm Hg120\ \text{mm Hg} and diastolic pressure is below 80 mm Hg80\ \text{mm Hg}.

  • Elevated: systolic pressure is 120–129 mm Hg120\text{–}129\ \text{mm Hg} and diastolic pressure is below 80 mm Hg80\ \text{mm Hg}.

  • Hypertension, stage 1: systolic pressure is 130–139 mm Hg130\text{–}139\ \text{mm Hg} or diastolic pressure is 80–89 mm Hg80\text{–}89\ \text{mm Hg}.

  • Hypertension, stage 2: systolic pressure is at least 140 mm Hg140\ \text{mm Hg} or diastolic pressure is at least 90 mm Hg90\ \text{mm Hg}.

These adult categories do not replace age- and condition-specific assessment. A hypertension diagnosis should not be based on one reading alone. For example, a reading of 134/86 mm Hg134/86\ \text{mm Hg} falls in stage 1 because either its systolic or diastolic value meets that category; diagnosis still requires appropriately repeated readings and clinical assessment.

Very high readings

For an adult with a reading above 180 mm Hg180\ \text{mm Hg} systolic and/or 120 mm Hg120\ \text{mm Hg} diastolic, repeat the measurement after at least 11 minute. If it remains this high, contact a health professional promptly. If the reading is accompanied by chest pain, shortness of breath, weakness or numbness, vision change, difficulty speaking, or other concerning acute symptoms, call emergency services.

Recognizing deterioration

Assess the person, not just the monitor. A falling or MAP, a new marked rise in heart rate, altered mental status, fainting, cool or mottled skin, weak pulses, reduced urine output, or increasing breathing difficulty can signal worsening perfusion or another acute problem.

A normal does not rule out early deterioration. Compensatory responses can temporarily preserve pressure, and trends may reveal change before a single reading does. Recheck an unexpected measurement, assess symptoms and other vital signs, compare with the person's baseline, and promptly escalate concerning findings according to local protocols.

For example, a person with suspected blood loss may have near-normal while becoming increasingly confused, tachycardic, and cool to the touch. Do not assume the person is stable on the basis of the ; reassess promptly and escalate care.

Worked examples and core relationships

A pressure of 126/78 mm Hg126/78\ \text{mm Hg} gives an estimated MAP of about 94 mm Hg94\ \text{mm Hg}:

MAP≈126+2×783=94 mm HgMAP \approx \frac{126 + 2 \times 78}{3} = 94\ \text{mm Hg}

With a heart rate of 8080 beats per minute and of 65 mL/beat65\ \text{mL/beat}, is 5,200 mL/min5{,}200\ \text{mL/min}, or 5.2 L/min5.2\ \text{L/min}:

CO=80×65=5,200 mL/minCO = 80 \times 65 = 5{,}200\ \text{mL/min}

A cuff that is too small may read falsely high. A transducer below the level of the right atrium may display a falsely high pressure; re-level it and confirm the system according to protocol. A transducer positioned too high may display a falsely low pressure.

reflects the interaction of and vascular resistance. depends on heart rate and , while , contractility, and affect . Accurate interpretation depends on proper measurement technique, repeated readings and trends, and consideration of symptoms, other vital signs, and perfusion findings.